Introduction
PLEASE NOTE: This repair guide was developed by the iFixit team based on Welch Allyn’s own service manual. Neither iFixit nor this repair guide is endorsed by or affiliated with Welch Allyn.
What you need
Step 1
Stand
- Unplug the AC power connection cable.
Unplug the AC power connection cable.
1024
Step 2
- Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.
- Always grasp the hose by the connector. Do not pull on the hose itself.
- Pull the connector away from the connector port.
Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.
Always grasp the hose by the connector. Do not pull on the hose itself.
Pull the connector away from the connector port.
Step 3
- Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.
- Always grasp the cable by the connector. Do not pull on the cable itself.
- Pull the connector away from the connector port.
Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.
Always grasp the cable by the connector. Do not pull on the cable itself.
Step 4
- Grasp the temperature probe and pull it up to remove it from the monitor.
Grasp the temperature probe and pull it up to remove it from the monitor.
Step 5
- Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.
Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.
Step 6
- Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.
Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.
Step 7
- Remove the flathead screw on the USB networking door.
- Loosen the captive Phillips #2 screw securing the monitor to the stand.
Remove the flathead screw on the USB networking door.
Loosen the captive Phillips #2 screw securing the monitor to the stand.
Step 8
- Holding the monitor securely, open the USB networking door.
- You may have to lift the monitor slightly to allow the door to swing out from the stand mount.
Holding the monitor securely, open the USB networking door.
You may have to lift the monitor slightly to allow the door to swing out from the stand mount.
Step 9
- Detach any accessory USB cables from USB ports on the monitor.
Detach any accessory USB cables from USB ports on the monitor.
Step 10
- Slide the monitor off of the stand tray.
Slide the monitor off of the stand tray.
Step 11
Battery
- Place the monitor upside down on a clean surface.
- Insert a coin into the slot and push to open.
- Choose a coin that fits comfortably into the slot.
- Remove the battery cover.
Place the monitor upside down on a clean surface.
Insert a coin into the slot and push to open.
Choose a coin that fits comfortably into the slot.
Remove the battery cover.
Step 12
- Use the plastic label to remove the battery from its recess.
Use the plastic label to remove the battery from its recess.
To reassemble your device, follow these instructions in reverse order.