Introduction
Removing the Exhalation Corner Assembly is discussed in service manual revision D:
- Chapter 4: Assembly and Disassembly, Pg 51-52
PLEASE NOTE: This repair guide was developed by a student iFixit team based on Viasys/CareFusion’s own service manual. Neither iFixit nor this repair guide is endorsed by or affiliated with Viasys or CareFusion.
Step 1
Viasys Avea Exhalation Corner Assembly Removal
- Check to see if the exhalation filter assembly is installed, if it is, remove the exhalation filter assembly.
- Rotate the metal locking lever on the lower right of the ventilator body forward to an open position.
- Remove the exhalation filter assembly from the ventilator body by pulling straight down.
Check to see if the exhalation filter assembly is installed, if it is, remove the exhalation filter assembly.
Rotate the metal locking lever on the lower right of the ventilator body forward to an open position.
Remove the exhalation filter assembly from the ventilator body by pulling straight down.
1024
Step 2
- Using a long Phillips screwdriver, remove the Phillips screw located at the top of the exhalation filter assembly well, if applicable.
Using a long Phillips screwdriver, remove the Phillips screw located at the top of the exhalation filter assembly well, if applicable.
Step 3
- Using a long Phillips screwdriver, remove the Phillips screw located at the exhalation port marked EXH, if applicable.
Using a long Phillips screwdriver, remove the Phillips screw located at the exhalation port marked EXH, if applicable.
Step 4
- Using a long Phillips screwdriver, remove the Phillips screw from the outside of the plastic cover.
Using a long Phillips screwdriver, remove the Phillips screw from the outside of the plastic cover.
Step 5
- Using a long Phillips screwdriver, remove the Phillips screw on the font left wall.
- Using a long Phillips screwdriver, remove the Phillips screw on the right side of the wall.
- Set the plastic cover aside.
Using a long Phillips screwdriver, remove the Phillips screw on the font left wall.
Using a long Phillips screwdriver, remove the Phillips screw on the right side of the wall.
Set the plastic cover aside.
Step 6
- Using a long Phillips screwdriver, remove the 3 Phillips screws on the metal re-enforcement cover, if applicable, and set aside.
- Remove the Exhalation Corner Assembly.
Using a long Phillips screwdriver, remove the 3 Phillips screws on the metal re-enforcement cover, if applicable, and set aside.
Remove the Exhalation Corner Assembly.
To reassemble your device, follow the steps in the reverse order.
Cancel: I did not complete this guide.
One other person completed this guide.
Author
with 1 other contributor
Alden Glassey
Member since: 04/19/2018
486 Reputation
3 Guides authored
Badges:
11
+8 more badges
Team
Cal Poly, Team S1-G1, Zimmerman Spring 2020
Member of Cal Poly, Team S1-G1, Zimmerman Spring 2020
CPSU-ZIMMERMAN-S20S1G1
5 Members
6 Guides authored