Difficulty
Moderate
Steps
10
Time Required
15 minutes
Sections
2
- Access to Housing in the Ulco Integrus PSV
- 5 steps
- Battery
- 5 steps
Flags
1
In Progress
This guide is a work in progress. Reload periodically to see the latest changes!
BackUlco Integrus PSV Ventilator
Full Screen
Options
History
Save to Favorites
Download PDF
Edit
Translate
Get Shareable Link
Embed This Guide
Notify Me of Changes
Stop Notifications
Introduction
- Make sure anaesthetic machine is OFFDisconnect the machine from power supplyGain access to ventilator housing from guide below
What you need
Step 1
Access to Housing in the Ulco Integrus PSV
- Before disassembly, make sure the front wheel breaks are on.
Before disassembly, make sure the front wheel breaks are on.
1024
Step 2
- Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.
Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.
Step 3
- Follow the steps in this replacement guide.
Follow the steps in this replacement guide.
Step 4
- With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.
With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.
Step 5
- Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.
Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.
Step 6
Battery
- Remove the main cover plate from the ventilator enclosure by removing all screws.
Remove the main cover plate from the ventilator enclosure by removing all screws.
Step 7
- Disconnect the battery.
Disconnect the battery.
Step 8
- Remove the four pan head screws (SM0308P) holding down the battery plate (EV9109) and battery (EV9108) and remove the battery plate and battery.
Remove the four pan head screws (SM0308P) holding down the battery plate (EV9109) and battery (EV9108) and remove the battery plate and battery.
Step 9
- Install the new battery in the same location and screw down the battery plate onto the battery plate spacers so that the battery is firmly held in place.
Install the new battery in the same location and screw down the battery plate onto the battery plate spacers so that the battery is firmly held in place.
Step 10
- Reconnect the battery.
Reconnect the battery.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
Author
with 2 other contributors
Preston Myers
Member since: 04/09/2020
493 Reputation
4 Guides authored
Badges:
6
+3 more badges
Team
Cal Poly, Team S7-G13, Paton Spring 2020
Member of Cal Poly, Team S7-G13, Paton Spring 2020
CPSU-PATON-S20S7G13
3 Members
8 Guides authored