Introduction
- Make sure that the anesthetic machine is OFFGain access to ventilator housing from this guide here
What you need
Step 1
Access to Housing in the Ulco Integrus PSV
- Before disassembly, make sure the front wheel breaks are on.
Before disassembly, make sure the front wheel breaks are on.
1024
Step 2
- Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.
Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.
Step 3
- Follow the steps in this replacement guide.
Follow the steps in this replacement guide.
Step 4
- With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.
With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.
Step 5
- Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.
Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.
Step 6
O-Ring
- Pull out the exhaust gas tube (EV910424) and replace the O-ring in the groove on the tube and smear with silicone grease.
Pull out the exhaust gas tube (EV910424) and replace the O-ring in the groove on the tube and smear with silicone grease.
Step 7
- Pull out the drive gas tube (EV9104251) and replace the O-ring in the groove on the tube and smear with silicone grease.
Pull out the drive gas tube (EV9104251) and replace the O-ring in the groove on the tube and smear with silicone grease.
Step 8
- Pull out the blow off gas tube (EV910426).
Pull out the blow off gas tube (EV910426).
Step 9
- Unscrew the 12 retaining screws on the top plate. Carefully lever the top plate back taking care to ensure that none of the cables connecting it to the body of the ventilator become disconnected.
Unscrew the 12 retaining screws on the top plate. Carefully lever the top plate back taking care to ensure that none of the cables connecting it to the body of the ventilator become disconnected.
Step 10
- Using tweezers, remove the O-ring from inside the blow off gas tube receptacle on the flow block. Smear the O-ring groove with silicone grease and insert the new O-ring.
Using tweezers, remove the O-ring from inside the blow off gas tube receptacle on the flow block. Smear the O-ring groove with silicone grease and insert the new O-ring.
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
One other person completed this guide.
Author
with 2 other contributors
Preston Myers
Member since: 04/09/2020
493 Reputation
4 Guides authored
Badges:
6
+3 more badges
Team
Cal Poly, Team S7-G13, Paton Spring 2020
Member of Cal Poly, Team S7-G13, Paton Spring 2020
CPSU-PATON-S20S7G13
3 Members
8 Guides authored