Difficulty

Moderate

Steps

12

Time Required

                          15 minutes            

Sections

3

  • Access to Housing in the Ulco Integrus PSV
  • 5 steps
  • Removing the Rear Access Panels in the Ulco Integrus PSV
  • 2 steps
  • Battery
  • 5 steps

Flags

0

  • BackUlco Integrus PSV

  • Full Screen

  • Options

  • History

  • Save to Favorites

  • Download PDF

  • Edit

  • Translate

  • Get Shareable Link

  • Embed This Guide

  • Notify Me of Changes

  • Stop Notifications

Introduction

  • Make sure anaesthetic machine is OFFDisconnect the machine from power supplyGain access to ventilator housing from guide below

For more information on replacing the battery, please refer to page 29 in the service manual.

What you need

Step 1

              Access to Housing in the Ulco Integrus PSV               
  • Before disassembly, make sure the front wheel breaks are on.

Before disassembly, make sure the front wheel breaks are on.

1024

Step 2

  • Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.

Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.

Step 3

  • Follow the steps in this replacement guide.

Follow the steps in this replacement guide.

Step 4

  • With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.

With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.

Step 5

  • Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.

Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.

Step 6

              Removing the Rear Access Panels in the Ulco Integrus PSV               
  • Remove the rear plastic cover by turning the single hex key retaining clip to free the cover.

Remove the rear plastic cover by turning the single hex key retaining clip to free the cover.

Step 7

  • Remove the four screws mounted to the cover plate using a 4mm Allen key.

Remove the four screws mounted to the cover plate using a 4mm Allen key.

Step 8

              Battery               
  • Remove the main cover plate from the ventilator enclosure by removing all screws.

Remove the main cover plate from the ventilator enclosure by removing all screws.

Step 9

  • Disconnect the battery.

Disconnect the battery.

Step 10

  • Remove the four pan head screws (SM0308P) holding down the battery plate (EV9109) and battery (EV9108) and remove the battery plate and battery.

Remove the four pan head screws (SM0308P) holding down the battery plate (EV9109) and battery (EV9108) and remove the battery plate and battery.

Step 11

  • Install the new battery in the same location and screw down the battery plate onto the battery plate spacers so that the battery is firmly held in place.

Install the new battery in the same location and screw down the battery plate onto the battery plate spacers so that the battery is firmly held in place.

Step 12

  • Reconnect the battery.

Reconnect the battery.

To reassemble your device, follow these instructions in reverse order.

Cancel: I did not complete this guide.